Short answer
CLIA, 42 CFR 493.1252(b): "The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided. These conditions must be monitored and documented." Temperature is listed.
CAP COM.30750 (CAP 2025 top 10): storage devices and storage areas checked daily; temperature-dependent equipment and work areas checked each day of use; calibrated thermometer; written procedure.
D5413, the CMS tag for 493.1252(b) as a whole (temperature, humidity, water quality and power, not temperature alone), was the most cited standard-level deficiency in 2024: 818 laboratories (CMS).
What does CLIA require?
42 CFR 493.1252(b). Conditions to define, monitor and document, "if applicable": (1) water quality, (2) temperature, (3) humidity, (4) "Protection of equipment and instruments from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports."
Storage criteria not met: corrective action documented, 42 CFR 493.1282(b)(3).
What does CAP COM.30750 require?
- Storage devices: refrigerators, freezers, incubators.
- Equipment: water baths, heat blocks.
- Environments: ambient reagent or specimen storage; conditions for instrument operation and test performance.
- Frequency: daily for storage devices and storage environments. Each day of use for equipment and environments used at a specified temperature.
- Instrument: calibrated thermometer, per written procedure. Heat blocks and dry baths: thermocouple probe accepted.
- Recording: the number, or a mark on a graph matching a number. Manual entries identify the recorder; initials are enough.
- Exempt from the daily check: testing devices with built-in fail-safe that disables use when out of range.
- Frost-free freezers: specimens, reagents and controls only if protected from thawing (thermal containers, for example).
CAP's 2025 citation reasons for COM.30750 (CAP): missing days on manual logs; no corrective action for out-of-range temperatures; no min/max thermometer during closures in labs not open 24/7.
What about weekends, holidays and closures?
CAP COM.30750: a continuous monitoring device or a minimum/maximum thermometer satisfies daily recording during closures if the data is evaluated the next business day, before use. Min/max thermometer: reset before the closure; record low and high before using any temperature-dependent equipment, kits or supplies. Open days with daily temperatures recorded: low and high not needed.
What if a monitoring system records the temperatures?
Accepted by CAP in place of manual readings when (1) staff have ongoing immediate access to the data so they can act on an out-of-range temperature and (2) system records show daily functionality per the manufacturer's instructions. Routine daily review of system records: not required (CAP).
Different guidance for waived testing sites, CDC job aid: "If a continuous temperature monitoring system is in use, the daily temperature should still be reviewed." (CDC) For CAP-accredited laboratories, the CAP rule above applies.
Who sets the acceptable range?
The laboratory, consistent with the manufacturer's instructions (42 CFR 493.1252(b)). Typical ranges in the CDC job aid: refrigerator for patient samples 2 to 8 °C; sample freezer -25 to -15 °C; testing rooms 15 to 30 °C. Same source: "The proper temperature range for a freezer or refrigerator should be determined by considering the acceptable temperature range indicated for the reagents or test kits, controls, and patient samples that will be stored in it." (CDC) Check package inserts before adopting a template range.
What goes in an out-of-range entry?
Required: documented corrective action (42 CFR 493.1282(b)(3)). CDC job aid: stop testing, contact the person who directs or supervises testing, "Document all corrective actions." (CDC) Fields worth capturing:
- Reading, unit, date and time.
- Who found it.
- Action taken (door closed, service called, contents moved).
- Follow-up reading and time.
- Disposition of reagents, controls and specimens inside, and who decided.
- When the unit went back into use.
What if a reading was missed?
CDC job aid: "If a temperature reading is missed, and there is no available temperature data for that time frame, the entry log should remain blank. Do not make up or guess what the temperature was for that reading." (CDC) Add a note: why it was missed, what was checked after.
What should a month's temperature log hold?
- HeaderUnit, location, acceptable range, month.CDC job aid · 42 CFR 493.1252(b)
- Every required dayDaily for storage devices and storage areas; each day of use for equipment and work areas.CAP COM.30750
- Who recorded itInitials or name on each manual reading.CAP COM.30750
- ThermometerWhich calibrated thermometer or probe.CAP COM.30750
- ClosuresContinuous data, or min and max after a reset, evaluated next business day before use.CAP COM.30750
- Out-of-range readingsEach with corrective action and follow-up reading.42 CFR 493.1282(b)(3) · CAP COM.30750
- Missed readingsBlank, with a note.CDC job aid
- CorrectionsOriginal visible; who changed it, when.CAP GEN.20450
- ReviewSigned and dated at month end by the person who directs or supervises testing. This is the CDC job aid's practice for waived sites; follow your own procedure and accreditor.CDC job aid
- RetentionAt least 2 years.42 CFR 493.1105(a)(3)
Frequently asked questions
- How often do CAP laboratories record refrigerator temperatures?
- Daily, for refrigerators and freezers storing reagents, supplies or specimens within a specified range (CAP COM.30750).
- Do we need temperature records when the laboratory is closed?
- Yes. Continuous monitoring or a reset min/max thermometer covers the closure; the data is evaluated the next business day before use (CAP COM.30750).
- Is a continuous monitoring system enough on its own?
- CAP accepts it in place of manual readings when staff have immediate access to the data and system records show daily functionality. Routine daily review of system records is not required.
- Does a heat block need a daily reading?
- Equipment used at a specified temperature is checked each day of use. A thermocouple probe is accepted for heat blocks and dry baths (CAP COM.30750).
- How long do we keep temperature records?
- At least 2 years under CLIA, 42 CFR 493.1105(a)(3). Accreditors and states may require longer.
- Can we correct a temperature entry?
- Yes, with the original entry still visible or in an audit trail, plus who changed it and when (CAP GEN.20450). CAP names erasures and correction fluid or tape as unacceptable.
Sources
- 42 CFR 493.1252, Test systems, equipment, instruments, reagents, materials and supplieseCFR
- 42 CFR 493.1282, Corrective actionseCFR
- 42 CFR 493.1105, Retention requirementseCFR
- Year 2024 Top 10 Standard-Level DeficienciesCenters for Medicare & Medicaid Services
- 2025 CAP Top 10 Most Common Checklist Deficiencies (revised March 2026)College of American Pathologists, copy hosted by UT Health San Antonio
- Temperature Log Instructions job aidCenters for Disease Control and Prevention
This guide summarizes published requirements as of the review date. It is not legal or accreditation advice. Your accreditor's current checklist edition and your laboratory director govern how they apply to your laboratory.