Skip to content

Guide

Questions to ask when choosing laboratory operations software

Questions tied to what CAP and CLIA ask a laboratory to keep, and a demo script that makes a vendor show the record instead of describing it.

Last reviewed 6 minute read

Short answer

Ask to see records, live: who did the work and when, the original value after a correction, each task's frequency, the monthly review and its date, the corrective action for each out-of-range reading, and the export.

Same questions, same order, every vendor.

What should we do before the first demo?

  • List current records: maintenance, function checks, temperatures, monthly reviews, corrective actions, competency, controlled documents.
  • Count instruments, monitored units, sites.
  • Pull last inspection's findings. Use them as test cases.
  • Bring a bench technologist and a reviewer to the demo.
  • Get IT and privacy requirements before the shortlist.

Records and corrections

  1. After a correction, can we still see the original value, who changed it and when?

    Why ask CAP GEN.20450: "Original (erroneous) entries must be visible (ie, erasures and correction fluid or tape are unacceptable) or accessible (eg, audit trail for electronic records)." (CAP)

  2. Can anyone, admins and vendor staff included, delete or overwrite a record?

    Why ask CAP GEN.20450: "Corrected data, including the identity of the person changing the record and when the record was changed, must be accessible to audit." (CAP)

  3. Are person and time captured by the system, not typed?

    Why ask Our preference, not a CAP rule. CAP COM.30750: "If temperatures are recorded manually, the identity of the individual recording temperatures must be recorded (the person's initials are adequate)." (CAP)

Schedules and limits

  1. Does each task store its frequency and where that frequency came from?

    Why ask 42 CFR 493.1254(a)(1): "Maintenance as defined by the manufacturer and with at least the frequency specified by the manufacturer." Without a manufacturer protocol, CAP COM.30600: "the laboratory must establish a schedule and procedure that reasonably reflect the workload and operating specifications of its equipment."

  2. Who sets the acceptable range for a unit or function check? Does the software ever fill one in?

    Why ask 42 CFR 493.1252(b): "The laboratory must define criteria for those conditions that are essential for proper storage of reagents and specimens, accurate and reliable test system operation, and test result reporting. The criteria must be consistent with the manufacturer's instructions, if provided."

  3. What does the screen show the day after a task is missed?

    Why ask CAP's first listed reason for COM.30600 citations in 2025: "Records that are missing (daily, weekly, or monthly function check)" (CAP).

  4. Can equipment be marked out of service, with checks required before it returns?

    Why ask CAP COM.30600: "Equipment not in use does not require maintenance or function checks, but these checks must be performed before putting the equipment back into service."

Review and sign-off

  1. How does monthly review work? Is the reviewer and the date stored?

    Why ask CAP COM.04200: "Appropriate evidence of review includes both the reviewer's signature or initials and the review date."

  2. Is the review date shown next to the month it covers?

    Why ask A CAP COM.04200 citation reason: "Completed several months of review on the same date (not timely)."

  3. Does the review include less-used equipment?

    Why ask A CAP COM.04200 citation reason: "Not performed for less-frequently used equipment (eg, osmometer)."

  4. Can the director name a designee, and is that recorded?

    Why ask CAP COM.04200: "The laboratory director or designee reviews and assesses instrument and equipment maintenance and function check records at least monthly."

Out-of-range readings and corrective action

  1. Does an out-of-range reading open an item that stays open until an action is recorded?

    Why ask 42 CFR 493.1282(b): "The laboratory must document all corrective actions taken," including when "The criteria for proper storage of reagents and specimens, as specified under § 493.1252(b), are not met." CAP lists "Missing corrective action for out-of-range temperatures" among COM.30750 citation reasons.

  2. Can the action record what happened to affected reagents, controls and specimens?

    Why ask CDC job aid: "stop testing and contact the person who directs or supervises the testing for guidance and next steps, including if all impacted reagents or test kits should be discarded and when to resume testing. Document all corrective actions." (CDC)

  3. With sensors: immediate staff access to the data, and proof the system worked each day?

    Why ask CAP COM.30750: "laboratory personnel must have ongoing immediate access to the temperature data so that appropriate corrective action can be taken if a temperature is outside of the acceptable range. System records must demonstrate daily functionality of the automated system in accordance with manufacturer's instructions."

Data, privacy and leaving

  1. Does the software need patient information at all? If it holds PHI, will you sign a business associate agreement?

    Why ask HHS: a business associate creates, receives, maintains or transmits PHI on a covered entity's behalf (HHS). If the system holds no PHI, including in free-text notes, attachments and specimen identifiers, a BAA is generally not needed. Confirm with your privacy officer.

  2. Can we export every record, with history, in a format that opens without your software?

    Why ask 42 CFR 493.1105(a)(3): "Retain quality control and patient test records (including instrument printouts, if applicable) and records documenting all analytic systems activities specified in §§ 493.1252 through 493.1289 for at least 2 years." That clock keeps running after a contract ends.

  3. How long do you keep our records, and what happens if we leave or you shut down?

    Why ask 42 CFR 493.1105(b): "If the laboratory ceases operation, the laboratory must make provisions to ensure that all records and, as applicable, slides, blocks, and tissue are retained and available for the time frame specified in this section."

  4. Where are records stored, and who at your company can reach them?

    Why ask Your IT and privacy review will ask. Get it in writing.

How the software describes your status

  1. Is status shown as counts of open items that link to records, or as one score?

    Why ask Counts can be traced to records. A percentage can't.

  2. Does the software ever say we are compliant?

    Why ask That call belongs to the inspector or surveyor.

A demo script to send every vendor

Send ahead. Ask for each step live, in order.

  1. Record a reading outside the rangeWhat opens, who is notified, what closes it.42 CFR 493.1282(b)(3)
  2. Correct an earlier entryOriginal value, who, when, why.CAP GEN.20450
  3. Miss a scheduled taskHow it looks the next day and in monthly review.CAP COM.30600
  4. Sign a monthly reviewWhere reviewer and date appear; how a late review looks.CAP COM.04200
  5. Take an instrument out of serviceReturn it. Which checks are required first.CAP COM.30600
  6. Record at the benchYour technologist, your bench device, one maintenance task.CAP COM.30600
  7. Export one instrument's monthOpen the file without the vendor's software.42 CFR 493.1105

Frequently asked questions

Do we need separate software if our LIS has an equipment module?
Not necessarily. Ask the LIS vendor the same questions: history, dates, reviews.
Does CLIA require electronic records?
No. CAP COM.30600: "It is up to the laboratory to determine how and where records of maintenance checks are kept." CLIA sets retention at 2 years for analytic systems records (42 CFR 493.1105(a)(3)).
Does a software vendor need a business associate agreement?
Only if it creates, receives, maintains or transmits protected health information on your behalf (HHS). Ask whether the software needs patient information at all.
How long must the records be kept?
42 CFR 493.1105(a)(3): records documenting analytic systems activities "for at least 2 years." Accreditors and states may require longer.
Can software tell us we are compliant?
No. The inspector or surveyor decides. Prefer counts of open items linked to records over a single score.

Sources

  1. 42 CFR 493.1252, Test systems, equipment, instruments, reagents, materials and supplieseCFR
  2. 42 CFR 493.1254, Maintenance and function checkseCFR
  3. 42 CFR 493.1282, Corrective actionseCFR
  4. 42 CFR 493.1105, Retention requirementseCFR
  5. 2025 CAP Top 10 Most Common Checklist Deficiencies (revised March 2026)College of American Pathologists, copy hosted by UT Health San Antonio
  6. Temperature Log Instructions job aidCenters for Disease Control and Prevention
  7. Business Associates Frequently Asked QuestionsU.S. Department of Health and Human Services

This guide summarizes published requirements as of the review date. It is not legal or accreditation advice. Your accreditor's current checklist edition and your laboratory director govern how they apply to your laboratory.

Keep these records as the work happens

CLoMS records each reading, task, review and correction with the person, the time and the original value.