Short answer
CLIA, 42 CFR 493.1773(c): "A laboratory must have all records and data accessible and retrievable within a reasonable time frame during the course of the inspection."
CMS, 2024 (CMS): the most cited standard was 42 CFR 493.1252(b), "These conditions must be monitored and documented," at 818 laboratories. Then 493.1235 competency policies (778), 493.1251(a) procedure manual (635), 493.1251(b) manual content (609) and 493.1252(d) expired or deteriorated supplies (607).
CAP's 2025 top 10, as its document lists them (CAP): procedure manual (COM.10000), activity menu (COM.04075), proficiency testing evaluation (COM.01700), maintenance and function checks (COM.30600), competency (GEN.55500), instrument comparability (COM.04250), monthly equipment review (COM.04200), record corrections (GEN.20450), reagent labels (COM.30300), temperatures (COM.30750).
What can a surveyor require on the day?
42 CFR 493.1773(b): CMS or a CMS agent may require the laboratory to:
- "Test samples, including proficiency testing samples, or perform procedures."
- "Permit interviews of all personnel concerning the laboratory's compliance with the applicable requirements of this part."
- "Permit laboratory personnel to be observed performing all phases of the total testing process (preanalytic, analytic, and postanalytic)."
- Give access to specimen areas, testing and reporting areas, and "Storage facilities for specimens, reagents, supplies, records, and reports."
- "Provide CMS or a CMS agent with copies or exact duplicates of all records and data it requires."
Accredited laboratories too: a laboratory with a certificate of accreditation "must permit CMS or a CMS agent to conduct validation and complaint inspections" (42 CFR 493.1773(a)).
Procedure manual
- Available and followed: "A written procedure manual for all tests, assays, and examinations performed by the laboratory must be available to, and followed by, laboratory personnel" (42 CFR 493.1251(a)). CMS tag D5401, 635 laboratories in 2024.
- Approval: "Procedures and changes in procedures must be approved, signed, and dated by the current laboratory director before use" (42 CFR 493.1251(d)).
- History: each procedure kept "with the dates of initial use and discontinuance" (42 CFR 493.1251(e)), for at least 2 years after it is discontinued (42 CFR 493.1105(a)(2)).
- CAP COM.10000: "All policies and procedures, in either electronic or paper form, must be readily available for review by the inspector at the time of inspection by the CAP." Copies must also be reachable during network downtime.
CAP's 2025 citation reasons for COM.10000 (CAP): "Not following policy/procedure (eg, changed process but failed to update procedure)"; "Failure to archive a procedure not in use"; "Printed procedures don't match electronic documents (version control)".
Personnel and competency
- Written policy: "the laboratory must establish and follow written policies and procedures to assess employee and, if applicable, consultant competency" (42 CFR 493.1235). CMS tag D5209, 778 laboratories in 2024.
- Six elements for each person (42 CFR 493.1413(b)(8) for moderate complexity, 493.1451(b)(8) for high complexity): direct observation of routine testing; monitoring the recording and reporting of results; review of worksheets, QC, proficiency testing and preventive maintenance records; direct observation of maintenance and function checks; testing previously analyzed, blind or proficiency samples; problem solving.
- Frequency, the same for both: "at least semiannually during the first year the individual tests patient specimens. Thereafter, evaluations must be performed at least annually" (42 CFR 493.1413(b)(9), 493.1451(b)(9)), and again before reporting after a change in method or instrument.
- CAP GEN.55500: records hold all six elements "for each individual on each test system during each assessment period, unless an element is not applicable to the test system."
- Who may do what, in writing. Moderate complexity, 42 CFR 493.1407(e)(14): "Specify, in writing, the responsibilities and duties of each consultant and each person, engaged in the performance of the preanalytic, analytic, and postanalytic phases of testing". High complexity, 493.1445(e)(15): "each consultant and each supervisor, as well as each person engaged in the performance of the preanalytic, analytic, and postanalytic phases of testing". Both include "which examinations and procedures each individual is authorized to perform".
CAP's 2025 citation reasons for GEN.55500: "Not documenting all six elements that apply/elements assessed were unclear"; "Records supporting the assessment are not documented or available"; "Confusion about what is considered a test system (eg, lumping all manual tests under one test system)".
Proficiency testing
- The director ensures the laboratory is enrolled, samples are tested as required, results go back on time, "All proficiency testing reports received are reviewed by the appropriate staff", and "An approved corrective action plan is followed" for each unacceptable or unsatisfactory result (42 CFR 493.1407(e)(4), 493.1445(e)(4)).
- Tests not listed in subpart I, and listed tests with no compatible samples from a CMS-approved proficiency testing program: "At least twice annually, the laboratory must verify the accuracy" of each one (42 CFR 493.1236(c)). CMS tag D5217, 548 laboratories in 2024. "All proficiency testing evaluation and verification activities must be documented" (42 CFR 493.1236(d)).
- CAP COM.01700 evidence: records of review of every PT report by the director or designee, and of the investigation and corrective action for each unacceptable result. Primary records kept at least two years, five for transfusion medicine.
- Retention under CLIA: "Retain all proficiency testing records for at least 2 years" (42 CFR 493.1105(a)(4)).
Method verification and comparability
- Unmodified FDA-cleared or approved test system, before patient results (42 CFR 493.1253(b)(1)): "Demonstrate that it can obtain performance specifications comparable to those established by the manufacturer" for accuracy, precision and reportable range, and "Verify that the manufacturer's reference intervals (normal values) are appropriate for the laboratory's patient population." Under (c): "The laboratory must document all activities specified in this section." D5421: 451 labs in 2024.
- Kept "for the period of time the laboratory uses the test system but no less than 2 years" (42 CFR 493.1105(a)(3)(i)).
- CAP COM.04250: "If the laboratory uses more than one nonwaived instrument/method to test for a given analyte, the instruments and methods are checked against each other at least twice a calendar year for comparability of results." Cited in 2025 for "Lack of acceptance criteria" and "Not performed twice a year".
Equipment, temperatures and reagents
- Storage and operating conditions defined, "monitored and documented" (42 CFR 493.1252(b)). CMS tag D5413, the most cited standard in 2024: 818 laboratories.
- Maintenance and function checks at least as often as the manufacturer specifies, or under the laboratory's own protocol for equipment developed in-house, modified by the laboratory, or without manufacturer protocols (42 CFR 493.1254(a), (b)), reviewed by the director or designee at least monthly (CAP COM.04200).
- Temperatures: storage devices and environments "must be checked daily", and "If temperatures are recorded manually, the identity of the individual recording temperatures must be recorded" (CAP COM.30750).
- D5417, 42 CFR 493.1252(d): reagents and supplies "must not be used when they have exceeded their expiration date, have deteriorated, or are of substandard quality." 607 labs in 2024.
- CAP COM.30300 label elements: "Content and quantity, concentration or titer", "Storage requirements", "Date prepared, filtered or reconstituted by laboratory", "Expiration date". These "may be recorded in a log (paper or electronic), rather than on the containers themselves, providing that all containers are identified so they are traceable to the appropriate data in the log" (CAP).
Corrections and corrective action
- CAP GEN.20450: "Original (erroneous) entries must be visible (ie, erasures and correction fluid or tape are unacceptable) or accessible (eg, audit trail for electronic records)." Who changed a record and when must be "accessible to audit."
- CLIA: "The laboratory must document all corrective actions taken" (42 CFR 493.1282(b)).
A pull list for the week before
- Test menuCAP COM.04075: the activity menu "must include all patient/client testing performed by the laboratory."CAP COM.04075
- Procedures"approved, signed, and dated by the current laboratory director before use", each kept "with the dates of initial use and discontinuance".42 CFR 493.1251(d), (e) · CAP COM.10000
- PersonnelWritten duties naming "which examinations and procedures each individual is authorized to perform" (493.1407(e)(14), 493.1445(e)(15)). CAP GEN.55500: competency "for each individual on each test system during each assessment period".42 CFR 493.1407(e)(14) · 42 CFR 493.1445(e)(15) · 42 CFR 493.1413(b)(8), (9) · 42 CFR 493.1451(b)(8), (9) · CAP GEN.55500
- Proficiency testingCAP COM.01700: "Records of ongoing review of all PT reports and alternative performance assessment results by the laboratory director or designee" and "Records of investigation of each 'unacceptable' PT and alternative performance assessment result". CLIA: "All proficiency testing reports received are reviewed by the appropriate staff".CAP COM.01700 · 42 CFR 493.1407(e)(4) · 42 CFR 493.1445(e)(4)
- Tests outside approved PT"At least twice annually, the laboratory must verify the accuracy"; "All proficiency testing evaluation and verification activities must be documented."42 CFR 493.1236(c), (d)
- VerificationRetain "records of test system performance specifications that the laboratory establishes or verifies".42 CFR 493.1253(b) · 42 CFR 493.1105(a)(3)(i)
- Comparability"Records of comparability studies reflecting performance at least twice per year with appropriate specimen types".CAP COM.04250
- EquipmentMaintenance and function checks (493.1254). CAP COM.04200: "Records of monthly review" with "the reviewer's signature or initials and the review date".42 CFR 493.1254 · CAP COM.04200 · CAP COM.30600
- TemperaturesCAP COM.30750: "must be checked daily"; over closures, data "evaluated on the next business day prior to use". CLIA: conditions "monitored and documented".CAP COM.30750 · 42 CFR 493.1252(b)
- ReagentsCLIA 493.1252(c): labeled. CAP COM.30300 accepts "Properly labeled reagents OR logs traceable to the reagents".CAP COM.30300 · 42 CFR 493.1252(c)
- Corrections"Records of corrections to laboratory records", with who changed each and when "accessible to audit".CAP GEN.20450
Frequently asked questions
- How quickly do records have to be produced?
- 42 CFR 493.1773(c): "A laboratory must have all records and data accessible and retrievable within a reasonable time frame during the course of the inspection."
- Are electronic records acceptable?
- CAP COM.10000, note 5: "Electronic manuals accessed on the computer are fully acceptable. There is no requirement for paper copies to be available for the routine operation of the laboratory, as long as the electronic versions are readily available to all personnel and personnel have been trained on how to access them." CAP COM.30600: "It is up to the laboratory to determine how and where records of maintenance checks are kept."
- How long do these records have to be kept?
- 42 CFR 493.1105: analytic systems records "for at least 2 years"; "Retain all proficiency testing records for at least 2 years"; verification records "for the period of time the laboratory uses the test system but no less than 2 years"; "Pathology test reports for at least 10 years after the date of reporting." CAP COM.01700 keeps PT records "at least two years (five years for transfusion medicine)". States may set longer periods.
- We passed proficiency testing overall. Do the missed challenges still need review?
- Yes. CAP COM.01700: "Each unacceptable PT or alternative performance assessment result (any result or specimen not meeting defined acceptability criteria) must be evaluated in a timely manner." CAP's 2025 citation reasons include "Misunderstanding about the need to review missed challenges when the laboratory receives an overall passing score".
- Who has to sign a procedure?
- 42 CFR 493.1251(d): "Procedures and changes in procedures must be approved, signed, and dated by the current laboratory director before use."
- Can software tell us we are ready for an inspection?
- No. Under 42 CFR 493.1773(d) the laboratory provides "all information and data needed by CMS or a CMS agent to make a determination of the laboratory's compliance". On a CAP inspection, the inspector cites deficiencies against the checklist. CMS or the CAP makes the determination.
Sources
- 42 CFR 493.1773, Basic inspection requirements for all laboratories issued a CLIA certificate and CLIA-exempt laboratorieseCFR
- 42 CFR 493.1251, Procedure manualeCFR
- 42 CFR 493.1235, Personnel competency assessment policieseCFR
- 42 CFR 493.1413, Technical consultant responsibilitieseCFR
- 42 CFR 493.1451, Technical supervisor responsibilitieseCFR
- 42 CFR 493.1407, Laboratory director responsibilities (moderate complexity)eCFR
- 42 CFR 493.1445, Laboratory director responsibilities (high complexity)eCFR
- 42 CFR 493.1236, Evaluation of proficiency testing performanceeCFR
- 42 CFR 493.1253, Establishment and verification of performance specificationseCFR
- 42 CFR 493.1252, Test systems, equipment, instruments, reagents, materials and supplieseCFR
- 42 CFR 493.1254, Maintenance and function checkseCFR
- 42 CFR 493.1282, Corrective actionseCFR
- 42 CFR 493.1105, Retention requirementseCFR
- Year 2024 Top 10 Standard-Level DeficienciesCenters for Medicare & Medicaid Services
- 2025 CAP Top 10 Most Common Checklist Deficiencies (revised March 2026)College of American Pathologists, copy hosted by UT Health San Antonio
This guide summarizes published requirements as of the review date. It is not legal or accreditation advice. Your accreditor's current checklist edition and your laboratory director govern how they apply to your laboratory.