Short answer
CLIA, 42 CFR 493.1254(a): for unmodified manufacturer equipment, the laboratory must perform and document "Maintenance as defined by the manufacturer and with at least the frequency specified by the manufacturer" and "Function checks as defined by the manufacturer and with at least the frequency specified by the manufacturer. Function checks must be within the manufacturer's established limits before patient testing is conducted."
CAP COM.30600 (CAP 2025 top 10): "The laboratory performs and records appropriate maintenance and function checks for all instruments (eg, analyzers) and equipment (eg, centrifuges) following a defined schedule, at least as frequent as specified by the manufacturer." COM.04200: "The laboratory director or designee reviews and assesses instrument and equipment maintenance and function check records at least monthly."
Frequency of citation: CMS tag D5429 (maintenance at manufacturer frequency), 449 laboratories in 2024 (CMS). COM.30600 and COM.04200 both on CAP's 2025 top 10.
What does CLIA require?
42 CFR 493.1254, two cases.
- (a) Unmodified manufacturer equipment, instruments or test systems. Maintenance and function checks as the manufacturer defines them, at least as often as the manufacturer specifies. Function checks within the manufacturer's limits before patient testing.
- (b) Equipment developed in-house, modified by the laboratory, or with no manufacturer protocol. The laboratory must "Establish a maintenance protocol that ensures equipment, instrument, and test system performance that is necessary for accurate and reliable test results and test result reporting," define a function check protocol the same way, then perform and document both. Function checks within the laboratory's own limits before patient testing.
Survey tag: D5429. Ninth on the CMS list of standard-level deficiencies for 2024, 449 laboratories (CMS).
Equipment out of range: 42 CFR 493.1282(b)(1)(i) requires documented corrective action for "Equipment or methodologies that perform outside of established operating parameters or performance specifications."
What does CAP COM.30600 add?
- Scope: all instruments (analyzers) and all equipment. CAP's equipment example is the centrifuge.
- Schedule: defined, at least as frequent as the manufacturer's.
- No manufacturer instructions: the laboratory sets a schedule and procedure that reflect its workload and the equipment's operating specifications.
- What counts, in CAP's note: "Maintenance and function checks may include (but are not limited to) cleaning, electronic, mechanical, and operational checks. The purpose of a function check is to detect drift, instability, or malfunction, before the problem is allowed to affect test results."
- Where records live: the laboratory decides.
- Equipment not in use: no checks while idle, checks before it returns to service.
CAP's most common reasons for COM.30600 citations in 2025 (CAP): "Records that are missing (daily, weekly, or monthly function check)"; "Not following manufacturer's defined frequency"; "Employees not documenting performed maintenance".
What does the monthly review under COM.04200 need?
Who: laboratory director or designee. How often: at least monthly. Evidence: reviewer signature or initials and the review date. Problem found (for example, maintenance not done on schedule): reviewer records the corrective action. Tests under an individualized quality control plan (IQCP): the review also considers whether the risk assessment and QC plan need another look (CAP).
CAP's 2025 citation reasons for COM.04200:
- Missing review because of staffing, for example supervisors pulled to the bench.
- No review for less-used equipment. CAP's example: osmometer.
- Several months reviewed on the same date. CAP calls this not timely.
Catching up on a backlog: sign each month as it is reviewed, with that day's date, and record why the review is late and the corrective action. Do not date old reviews to the month they cover. A late review is still late.
The record checklist, instrument by instrument
One instrument's file, before an inspection or when designing a new log.
- Equipment listEvery instrument and piece of equipment in use, less-used ones included.CAP COM.30600 · CAP COM.04200
- ScheduleEach task, its frequency, and the source of that frequency: manufacturer instructions, or the laboratory's protocol if there are none.42 CFR 493.1254(a), (b) · CAP COM.30600
- Procedure and limitsHow each task is done. Limits each function check must meet before patient testing.42 CFR 493.1254(a)(2), (b)(2)
- Each task performedWhat, when, who, result. Function check result next to its limit.42 CFR 493.1254
- Missed or late tasksMarked missed or late, with corrective action. Never back-filled as on time.CAP COM.04200 · CAP GEN.20450
- Problems and repairsOut-of-parameter findings, corrective action, service visits, checks before return to use.42 CFR 493.1282(b)(1)(i)
- Out of serviceDates out of use. Checks done before return to service.CAP COM.30600
- Monthly reviewEvery instrument, every month. Signature or initials, date, any corrective action.CAP COM.04200
- CorrectionsOriginal entry visible or in an audit trail. Who changed it, when.CAP GEN.20450
- RetentionAt least 2 years.42 CFR 493.1105(a)(3)
How do maintenance records connect to competency assessment?
The six competency elements come from CLIA. For high complexity testing, 42 CFR 493.1451(b)(8) includes "(iii) Review of intermediate test results or worksheets, quality control records, proficiency testing results, and preventive maintenance records" and "(iv) Direct observation of performance of instrument maintenance and function checks."
CAP carries the same six elements in GEN.55500 and notes that element 3 can be assessed during the monthly supervisory review of those records (CAP).
How long do the records have to be kept?
42 CFR 493.1105(a)(3): retain "records documenting all analytic systems activities specified in §§ 493.1252 through 493.1289 for at least 2 years." Maintenance and function checks (493.1254) fall inside that range. Under 493.1105(b), a laboratory that ceases operation must still make provisions for the records to be retained. Accreditors and states may require longer; check your checklist edition.
Frequently asked questions
- Does a centrifuge need maintenance records?
- Yes. CAP COM.30600 covers all instruments and equipment; the centrifuge is CAP's own example.
- What if the manufacturer gives no maintenance schedule?
- The laboratory writes its own maintenance and function check protocol, with a schedule reflecting workload and operating specifications, then follows and documents it. 42 CFR 493.1254(b); CAP COM.30600.
- Can a supervisor sign several months of review at once?
- CAP lists several months reviewed on the same date as a common COM.04200 citation (not timely). Review each month soon after it closes.
- Does equipment that is not in use need maintenance records?
- CAP COM.30600: "Equipment not in use does not require maintenance or function checks, but these checks must be performed before putting the equipment back into service." Record the dates it was out of service.
- Who can perform the monthly review?
- The laboratory director or designee (CAP COM.04200). CAP's note: "Appropriate evidence of review includes both the reviewer's signature or initials and the review date."
- Can maintenance records be electronic?
- Yes. CAP leaves how and where records are kept to the laboratory. Electronic corrections keep the original entry accessible, with who changed it and when (CAP GEN.20450).
Sources
- 42 CFR 493.1254, Maintenance and function checkseCFR
- 42 CFR 493.1282, Corrective actionseCFR
- 42 CFR 493.1105, Retention requirementseCFR
- 42 CFR 493.1451, Technical supervisor responsibilitieseCFR
- Year 2024 Top 10 Standard-Level DeficienciesCenters for Medicare & Medicaid Services
- 2025 CAP Top 10 Most Common Checklist Deficiencies (revised March 2026)College of American Pathologists, copy hosted by UT Health San Antonio
This guide summarizes published requirements as of the review date. It is not legal or accreditation advice. Your accreditor's current checklist edition and your laboratory director govern how they apply to your laboratory.